Measuring Reef Aquarium Salinity: Density, SG and Calibration

When water evaporates from your reef aquarium, the salts remain and salinity rises. Replace evaporation losses with suitable fresh water, usually RO/DI water, rather than newly mixed saltwater. A water change is different: you remove salt as well as water, so replace that water with appropriately mixed saltwater. Regular measurements help you check that top-ups and water changes are keeping salinity stable.
This guide helps you identify what your instrument measures and investigate conflicting results before changing your aquarium water.
Salinity, density and SG: What does your instrument show?
- Salinity: Describes salt content, often displayed in the hobby as ppt or ‰. Record the unit printed on your scale.
- Density: Mass per volume, for example kg/l. The value depends on temperature.
- Specific gravity (SG): A ratio to the density of reference water, without a unit. The temperature convention used by the scale is part of the result.
Example: Seawater at 35 ‰ has a density of about 1.023 kg/l at 25 °C. A reading of 1.023 on an SG scale means something different. Record a result such as “35 ppt” or “1.023 kg/l at 25 °C”, rather than just a number. This makes comparisons between instruments useful.
Calibrating a refractometer: Follow its instructions
Before buying, check whether the instrument is designed for seawater or a pure salt solution. Their scales are not automatically interchangeable. Do not apply one conversion number from a forum to every model.
Clean the prism and cover before calibrating. Allow the instrument to reach the temperature specified in its instructions and use the required liquid: depending on the model, this may be pure water for setting zero or a suitable salinity standard. Wait the stated time, read the light–dark boundary and adjust as instructed. “ATC” compensates for temperature effects within the instrument’s specified range; it does not replace calibration or a clean prism.
Hydrometers and digital salinity testers
Let a floating hydrometer float freely in its cylinder without touching the wall or bottom. Remove attached air bubbles and check the water temperature, because the scale uses a specified temperature. Read at the point required by your model’s instructions: the water surface or the meniscus. Reading at a different point can change the result.
Calibrate a digital salinity tester with the solution specified for its measurement method. Rinse and care for the sensor according to its instructions, and wait for the reading to stabilise. A reference intended for a refractometer is not automatically suitable for a conductivity meter just because both say “35 ppt”.
Two instruments disagree: Our checking sequence
- Record the model, unit, reading and water temperature. Do not compare kg/l directly with SG.
- Check cleaning, calibration and the reference against each instrument's instructions.
- Measure a fresh sample again and record both results.
- If the difference persists, arrange an independent cross-check. Do not change the tank based only on an unexplained single reading.
Frequently asked questions
Is 1.023 too low for a reef aquarium?
It cannot be assessed without the unit and temperature. First establish whether the instrument displays density or SG.
Can I use any 35-ppt reference?
Check that the manufacturer and reference specify compatible measurement and calibration conditions. The number alone is insufficient.
Which instrument should I choose?
Choose a seawater instrument with instructions, a scale and a reference you can understand. A simple, repeatable routine matters more than another decimal place alone.